Measurement
Displaying Measurement Results
Measurements are shown on the display. If the upper judgment limit and lower judgment limit of resistance
or voltage are set, the judgment result is displayed.
When measurements are abnormal
If the measurement display is “---” or a negative value, the measurement is not normal because the test
leads are not connected correctly. Refer to the following, and check the test lead connections.
When the maximum value of the range is exceeded
The following table shows the maximum values of each resistance and voltage range. If the maximum
value is exceeded, the measurement is shown as “OVER”.
Judgment result
Measured
value
UP: The measured value is greater than
or equal to the upper judgment limit.
IN:
The measured value within the
upper and lower limits.
LOW: The measured value is less than or
equal to the lower judgment limit.
Resistance measure-
ment display
Voltage measure-
ment display
Possible cause
“---”
“---”
Poor contact to the DUT’s positive or negative terminal
“---”
This is normal behavior. Poor contact to the SOURCE terminal
Negative value
This is normal behavior. The SOURCE HIGH and LOW terminals are connected in reverse.
“---”
Negative value
The DUT’s positive and negative terminals are connected in reverse.
Range
Maximum value
Resistance measurement
3 mΩ
3.1 mΩ
30 mΩ
31 mΩ
300 mΩ
310 mΩ
3 Ω
3.1 Ω
Voltage measurement
(BIM1030)
6 V
6.3 V
60 V
63 V
300 V
315 V
Voltage measurement
(BIM1100)
6 V
6.3 V
60 V
63 V
600 V
630 V
1000 V
1050 V