Select Thermocouples and Start Profile
Include Thermocouple (TC) Labels (20 char. Max)
– Checking this will display a field below each TC
that is checked. Up to 20 characters can be used to describe the placement or location of that TC.
TC Number Check Box
– Place a check next to each TC channel that is going to be used for this profile.
TC number 1 is always used for the
AIR TC
. You must have at least one other TC being used.
Live Reading
– When the KICstart is on and plugged into the download cable you will see live
temperature updates for the TCs plugged into the
KICstart
. The TC checkbox must be checked as well to
get the temperature updates.
Battery Voltage
– This displays a live reading of the
KICstart
battery voltage. The software will
recognize if your voltage is too low and not allow you to start a profile until you have a battery with enough
voltage to complete the profile.
KICstart Internal Temperature
– This displays the internal temperature of the
KICstart
profiler. If the
temperature is too high to complete the profile and stay under the maximum temperature rating for the unit
(105C), it will not allow you to start the profile. You must wait until the
KICstart
is cool enough. The
software will let you know what temperature you will have to drop below.
Buttons
Back (Arrow)
-This button takes you back to the previous screen.
Start Profile (Green Traffic Light)
- Select this button to start the profile.
Cancel (Red X)
-This button cancels the profile.