3
KICstart Hardware
Inventory
Part Description
KIC Part Number
KICstart Data logger (6 channel)
KICSTART-06
Communication Cable
CB-RS232-15
Thermal Shield
TS-SL-18
Individual Color Coded Thermocouples
TC-TK-06-36
Aluminum Tape and Hi-temp Solder for attaching Thermocouples
TAPE
KICstart User Manual, KICstart software
MAN-KS-V2
KICstart Configuration
The
KIC
start
– Datalogger
will be plugged in directly to the
computer COM port using the communication cable.
KICstart Sample rates
KICstart PROM Version 7.0
Application Type
Samples/Sec
Solder Reflow
4
Wave Solder
10
Cure
3
Semiconductor
4
The
KIC
start sample rates will vary depending on the selected
Application
type.
KICstart Maintenance: Changing the Battery
To install or change the battery or for calibration of the
KICstart
profiler you will have to unscrew the
battery cover from the bottom side of the
KICstart
. This will require a small Flathead screwdriver. The
screw has a retaining clip to keep it in place so it will not come all the way out. Just loosen it enough so the
cover comes free. Remove the battery cover. There is a white connector for a 9-volt battery. Once you
have connected the battery, slip the battery back in, placing the connector end in first. Then replace the
cover and tighten the battery cover. See the
KICstart
hardware section of this manual for calibration
details,
page 43.