![Keysight VXI Bus 75000 C Series Service Manual Download Page 32](http://html1.mh-extra.com/html/keysight/vxi-bus-75000-c-series/vxi-bus-75000-c-series_service-manual_1961145032.webp)
34
Keysight E1463A 32-Channel Form C Switch Service Guide
Verification Tests
Test 2-2: DC Isolation Test
NC (&C) to Chassis Isolation
1
Make hardware connections as shown in Figure 2-5
2
Set DMM to 2-wire ohms, 1 G
range
3
Send *RST to Module
4
Trigger the DMM with TRIG SGL
5
Record the DMM reading in Table 2-1 (NC to Chassis)
Figure 1-8
NO and NC to Chassis Isolation