![Keysight U8903B User Manual Download Page 577](http://html.mh-extra.com/html/keysight/u8903b/u8903b_user-manual_1961457577.webp)
Appendixes
A
Keysight U8903B User’s Guide
577
Example 4: Running the Test Sequence App for the multitone analyzer
measurement
In this example, you will learn how to use the Test Sequence App for the multitone analyzer
measurement in the U8903B.
1
Press
to switch to Test Sequence App mode.
2
In the Project tab, press
Add Measurement
, and select
Multitone Analyzer
to be added for
Test 1.
3
Press
Settings
to configure the multitone analyzer. For example, you can press
Waveform
Config
to set the amplitude, and press
Tones Config
to set the start and stop frequency.
4
The multitone measurement results can be displayed in a graph (Spectrum, Waveform,
Level, and Gain) and in a bar chart (Max Tone Level, Min Tone Level, TD+N Level, TD+N
Ratio, and Tone Level). Refer to “
information.
5
Press
and select
Start Test App
to begin the test sequence.
6
Depending on the settings parameter set for the multitone analyzer, the result will
return as PASS or FAIL and is shown as follows.
Figure A-14
Test Sequence App result for multitone analyzer
Summary of Contents for U8903B
Page 1: ...Keysight U8903B Performance Audio Analyzer User s Guide ...
Page 16: ...16 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 32: ...32 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 278: ...5 Graph Analysis 278 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 290: ...6 Sweep Function 290 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...