3
Performance Verification for Analog Analyzer
50
Keysight U8903B Service Guide
16
Compare the recorded CMRR values to the specification. If the test fails, adjustment will
need to be carried out accordingly.
Distortion (THD+N) verification
This test verifies if the residual THD+N of the U8903B is within the published specification
using an internal loopback connection to test the THD+N performance.
1
Preset the U8903B audio analyzer. Configure the U8903B generator output channel 1
and 2 to connect to all of its own analyzer input channels (up to 8 channels
concurrently) through internal loopback.
2
Configure the U8903B generator output channels as follows:
– Connection type: Unbalanced
– Output impedance: 600
Ω
– Output grounding configuration: Floating ground
– Waveform: Sine
– Output frequency: 1 kHz
– Output amplitude: 1 Vrms
– Output state: ON
3
Configure the U8903B analyzer input channels to be tested as follows:
– Connection type: Loopback
– Input bandwidth: 90 kHz
– Input coupling: DC
– Input range:1 V
– Voltage detector type: RMS
– Trigger source: Immediate
– Measurement function1: Frequency
– Measurement function2: THD+N Ratio
Table 3-7
Distortion specification
Distortion specification (at 1 kHz, 1 Vrms)
Equipment used
≤
–108 dB (for 20 Hz to 20 kHz bandwidth, 23 °C ±5 °C)
≤
–100 dB (for 20 Hz to 20 kHz bandwidth, 0 °C to 55 °C)
U8903B (internal loopback)
NOTE
Power on the U8903B audio analyzer for at least 30 minutes before performing the
performance verification.
Summary of Contents for U8903B
Page 1: ...Keysight U8903B Performance Audio Analyzer Service Guide ...
Page 12: ...12 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 126: ...A Appendix A 126 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 160: ...B Appendix B 160 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 164: ...C Appendix C 164 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...