The U8903A generator output settings can be configured using selectable output
connections, impedances, and reference impedances as shown in the figure below.
Figure 3-17
Output settings
Output type
The output connection can be set to Balanced, Unbalanced, or Common mode.
Balanced
outputs a pair of differential signals which are equal in amplitude but
180
°
out of phase on the XLR positive and negative pins.
Unbalanced
outputs a signal referenced to ground on the BNC output connector.
Common mode
outputs a pair of equal amplitude and in-phase signals on the XLR
positive and negative pins.