![Keysight U4421A Hardware And Probing Manual Download Page 41](http://html1.mh-extra.com/html/keysight/u4421a/u4421a_hardware-and-probing-manual_1961143041.webp)
Keysight U4421A MIPI D-PHY Analyzer and Exerciser Hardware and Probing Guide
41
Setting up an E5405A Soft Touch Midbus Probe Configuration for U4421A Module
5
Probe Pad Layout / Footprint Dimensions
The retention module alignment is symmetrical around the pad footprint.
Figure 19
Top view footprint dimensions (drawing notes below).
Drawing notes:
1 Maintain a solder mask web between pads when traces are routed between the pads on the same
layer. The solder mask may not encroach onto the pads within the pad dimension shown.
2 VIAs not allowed on these pads. VIA edges may be tangent to pad edges as long as a solder mask
web between VIAs and pads is maintained.
A1
A27
B1
B27
A1
A27
B1
B27
NOTE
The above view is looking down onto the footprint on the printed-circuit board.