5
Characteristics and Specifications
DC Power Supply Specifications
178
U3606B User’s Guide
Supplemental characteristics
Table 5-9
DC power supply supplemental characteristics
Parameter
Characteristics
S1
S1m
S2
S2m
Output ratings
30 V/1 A
100 mA/30 V
8 V/3 A
1000 mV/3 A
Maximum output
programming range
Voltage
CV: 31.500 V
OC: 1.05 A
OCP: 1.1 A
-
CV: 8.4 V
OC: 3.15 A
OCP: 3.3 A
CV: 1050 mV
OC: 3.15 A
OCP: 3.3 A
Current
CC: 1.05 A
OV: 31.500 V
OCP: 33.000 V
CC: 105 mA
OV: 31.500 V
OCP: 33.000 V
CC: 3.15 A
OV: 8.4 V
OCP: 8.8 V
-
Temperature coefficient
± (% of offset)
maximum change in output
/readback per °C for 0 ºC to 18
ºC/28ºC to 55 ºC
Voltage
0.005%
+ 0.5 mV
-
0.005%
+ 0.5 mV
0.005%
+ 0.05 mV
Current
0.02% +1 mA
0.02% +0.01
mA
0.02% +1 mA
-
Remote sensing capability
Voltage drop
per load lead
Up to 0.75 V
Load
regulation
< 0.01% + 3
mV
< 0.01% + 3
mV
< 0.01% + 3
mV
< 0.01%
+ 0.3 mV
Maximum load
voltage
Subtract voltage drop per load lead
Voltage programming speed
(excludes command processing
time)
Full load
Up: 50 ms
Down: 50 ms
No load
Up: 50 ms
Down: 50 ms
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply User s Guide ...
Page 2: ...THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 3: ...U3606B User s Guide I ...
Page 16: ...XIV U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 25: ...Introduction 1 Preparing the U3606B U3606B User s Guide 7 Adjust the carry handle 1 2a 3 2b ...
Page 27: ...Introduction 1 Preparing the U3606B U3606B User s Guide 9 Figure 1 3 Rack mount dimensions ...
Page 120: ...3 DC Power Supply Operation Square Wave Operation 102 U3606B User s Guide Enable the output 7 ...