Calibration Procedures
80
Keysight U3606B Service Guide
AC voltage gain adjustment procedure
Review the
on page 25 and
on page 77 sections before beginning this procedure.
1
Press
[
]
twice to enter the AC voltage gain calibration. The AC annunciator
will turn on according to your selection.
2
The display shows the uncalibrated value and the reference value of the
adjustment item.
3
Configure each adjustment item shown in
.
4
Use
or
to select the adjustment item.
5
Apply the input signal shown in the “Input” column of
6
Enter the actual applied input.
10.0000 V
10 V
100.000 V
100 V
1000.00 V
1000 V
Table 1-24
DC voltage gain adjustment (continued)
Adjustment item
Input voltage
CAUTION
Set the calibrator output to 0 V before disconnecting from the instrument
input terminals.
V
NOTE
Always complete tests in the same order as shown in