Calibration Procedures
1
Keysight U3606B Service Guide
41
Table 1-2
Zero offset verification test
Input
Function
Range
Error from nominal 1-year
Short
DC voltage
19.9999 mV
±0.008 mV
100.000 mV
±0.008 mV
1.00000 V
±0.00005 V
10.0000 V
±0.0005 V
100.000 V
±0.005 V
1000.00 V
±0.05 V
2-wire resistance
100
Ω
±0.008
Ω
1000
Ω
±0.05
Ω
10 k
Ω
±0.0005 k
Ω
100 k
Ω
±0.005 k
Ω
1 M
Ω
±0.00005 M
Ω
10 M
Ω
±0.0005 M
Ω
100 M
Ω
±0.005 M
Ω
4-wire resistance
0.1
Ω
±0.00005
Ω
1
Ω
±0.0003
Ω
10
Ω
±0.003
Ω
100
Ω
±0.03
Ω
1000
Ω
±0.3
Ω
CAUTION
Zero offset calibration using a multifunction calibrator is NOT recommended.
The calibrator and cabling offset can be large and unstable causing poor offset
calibration of the Keysight U3606B or any multimeter.