1
Calibration Procedures
38
Keysight U3606B Service Guide
Performance Verification Tests Overview
Use the verification and performance tests to verify the measurement
performance of the instrument. The verification and performance tests use the
specifications of the instrument listed in
Chapter 5 of the U3606B User’s Guide,
“Characteristics and Specifications,” starting on page 155
.
You can perform three different levels of verification and performance tests:
Self-tests
A series of internal verification tests, which gives a high confidence
that the instrument is operational.
Performance verification tests
An extensive set of tests that are
recommended as an acceptance test when you first receive the instrument
or after performing adjustments.
Additional verification tests
Tests that are not performed with every
calibration. Perform these tests to verify additional specifications or functions
of the instrument.