![Keysight U3606A User'S And Service Manual Download Page 165](http://html.mh-extra.com/html/keysight/u3606a/u3606a_users-and-service-manual_1961516165.webp)
Verification and Performance Tests
5
Recommended Test Equipment
U3606A User’s and Service Guide
143
General measurement techniques
A measurement made across the load includes the impedance of the leads
to the load. The impedance of the load leads can easily be several orders
of magnitude greater than the instrument impedance and thus invalidate
the measurement. To avoid mutual coupling effects, each measuring device
must be connected directly to the output terminals by using a separate
pair of leads.
Using an electronic load
Many of the output test verification procedures require the use of a
variable load resistor capable of dissipating the required power. Using a
variable load resistor requires that switches should be used to connect,
disconnect, and short the load resistor. An electronic load however, if
available, can be used in place of a variable load resistor. The electronic
load is considerably easier to use than load resistors. It eliminates the
need for connecting resistors or rheostats in parallel to handle power, it is
much more stable than a carbon- pile load, and it simplifies the work of
switching between load conditions as is required for the load regulation
and load response tests. Substitution of the electronic load requires minor
changes to the test procedures in this chapter.
Square wave
Oscilloscope
Keysight
DSO8064A or
Infiniium
equivalent
•
Sensitivity: 1 mV
•
Bandwidth limit: 20 MHz
•
Probe: 1:1 with RF tip; 10:1 with RF tip for > 50 V
Universal
frequency counter
Keysight 53131A
-
Table 5-1
Recommended test equipment for performance verification procedures (continued)
Application
Recommended equipment
Recommended
accuracy requirements
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...