U3606A Multimeter|DC Power Supply
Programmer’s Reference
4
CONFigure Subsystem
CONFigure[:VOLTage]:ACDC|DCAC 87
CONFigure:CURRent:ACDC|DCAC 96
This chapter describes the CONFigure commands used to program the
U3606A over a remote interface. Use the CONFigure commands to set the
measurement function, range, and resolution without actually making a
measurement.
N O T E
•
Use the
INITiate[:IMMediate]
or
READ?
command to initiate the
measurement.
•
Some measurements may result in a delayed response time in the
U3606A Multimeter|DC Power Supply. It is recommended that you
increase the SCPI query timeout to 15000 milliseconds or longer to
avoid SCPI query timeout errors.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Page 2: ......
Page 3: ......
Page 14: ...XII U3606A Programmer s Reference ...
Page 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Page 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Page 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Page 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Page 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Page 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Page 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Page 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Page 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Page 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Page 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...