CALCulate Subsystem
2
CALCulate[:STATe]
U3606A Programmer’s Reference
23
CALCulate[:STATe]
Syntax
CALCulate[:STATe] {0|1|OFF|ON}
This command turns the CALCulate subsystem, and thus the selected
calculation function, on or off.
CALCulate[:STATe]?
This query returns a boolean value that represents the current calculation
state:
0
or
1
Parameter
Remarks
•
This is an adjunct command to the
CALCulate:FUNCtion
command.
The calculation function to be used is selected using the
CALCulate:FUNCtion
command.
•
The
CALCulate:STATe
is set to
OFF
when the measurement function
is changed.
•
When the
CALCulate:STATe ON
command is sent, the math registers
for null, averaging, and dB reference value are cleared. This also occurs
when the
CALCulate:FUNCtion
command is sent with the
CALCulate:STATe
previously set to
ON
. The dBm reference resistance
value is not cleared in either case.
•
The instrument resets the calculation state to off after a Factory Reset
(
*RST
command), an Instrument Preset (
SYSTem:PRESet
command), or
a function change.
Item
Type
Range of values
Default value
state
Boolean
0|1|OFF|ON
0
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Page 2: ......
Page 3: ......
Page 14: ...XII U3606A Programmer s Reference ...
Page 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Page 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Page 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Page 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Page 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Page 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Page 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Page 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Page 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Page 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Page 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...