SENSe Subsystem
9
[SENSe:]DCYCle:VOLTage:RANGe[:UPPer]
U3606A Programmer’s Reference
231
[SENSe:]DCYCle:VOLTage:RANGe[:UPPer]
Syntax
[SENSe:]DCYCle:VOLTage:RANGe[:UPPer]
{<voltage_range>|MIN|MAX|DEF}
This command selects the measurement range for duty cycle measurements
via the voltage path.
[SENSe:]DCYCle:VOLTage:RANGe[:UPPer]? [{MIN|MAX}]
This query returns a numeric value that represents the duty cycle
measurement range via the voltage path.
Parameter
Remarks
•
Selecting a discrete range will disable autoranging.
•
If the input signal is greater than can be measured on the selected
range (manual ranging), the instrument gives an overload indication.
•
Positive overload: “
+9.37
” or “OL” from the front panel.
•
Negative overload: “
-9.37
” or “–OL” from the front panel.
N O T E
The
[SENSe:]DCYCle:VOLTage:RANGe[:UPPer]?
query is not
applicable when autoranging is enabled. It is recommended to manually
select the voltage range for frequency, pulse width, and duty cycle
measurements.
Item
Type
Range of values
Default value
voltage_
range
Numeric
•
100 mV|MIN
•
1 V
•
10 V
•
100 V
•
750 V|MAX
•
DEF (AUTO)
MIN
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Page 2: ......
Page 3: ......
Page 14: ...XII U3606A Programmer s Reference ...
Page 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Page 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Page 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Page 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Page 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Page 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Page 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Page 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Page 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Page 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Page 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...