6
MEASure Subsystem
MEASure[:VOLTage]:AC?
128
U3606A Programmer’s Reference
MEASure[:VOLTage]:AC?
Syntax
MEASure[:VOLTage]:AC?
[<range>|AUTO|MAX|MIN|DEF[,{<resolution>|MAX|MIN|DEF}]]
This command first resets all AC voltage measurement parameters and
trigger parameters to their default values. Then, it configures the
instrument for AC voltage measurements and immediately triggers a
measurement. A single reading is returned.
Parameters
C A U T I O N
•
The maximum range parameter is 750 V
rms
, which is set by MAX.
The rms voltage is waveform dependent. A sine wave is limited to
750 V
ac
(rms), but a 1000 V
pk
square wave is safe. Connections to
AC MAINS are further limited to CAT II (300V).
•
See the “Safety Information” section in the
U3606A User's and
Service Guide
for a complete discussion of the safety features, and
the procedures for safe operation of this instrument.
Item
Type
Range of values
Default value
range
Numeric
•
100 mV|MIN
•
1 V
•
10 V
•
100 V
•
750 V|MAX
•
AUTO
AUTO
resolution
Numeric
•
MAX (4½ digit)
•
10 uV
•
100 uV
•
1 mV
•
10 mV
•
100 mV
•
MIN (5½ digit)
•
1 uV
•
10 uV
•
100 uV
•
1 mV
•
10 mV
MIN
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Page 2: ......
Page 3: ......
Page 14: ...XII U3606A Programmer s Reference ...
Page 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Page 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Page 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Page 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Page 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Page 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Page 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Page 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Page 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Page 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Page 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...