![Keysight U2741A Programmer'S Reference Manual Download Page 28](http://html.mh-extra.com/html/keysight/u2741a/u2741a_programmers-reference-manual_1961514028.webp)
3
CALibrate Subsystem
28
Keysight U2741A Programmer’s Reference Guide
CALibration:ADC?
Syntax
CALibration:ADC?
This command performs a low-level calibration of the analog-to-digital (ADC)
converter circuitry. This is the first step in a recalibration sequence, and must be
done before doing the offset calibration.
Remarks
It will return a
+0
if the calibration passes and
+1
if it fails. Refer to the
Keysight
U2741A USB Modular 5.5 Digits Digital Multimeter Service Guide
for detailed
procedures, including how to connect to a calibration source, recommended
equipments, the specified calibration points, and so forth.
Example
CAL:ADC?
//Commands the U2741A to execute the
calibration
Typical Response:
+0
See Also