SENSe Subsystem
6
Keysight U2741A Programmer’s Reference Guide
107
[SENSe:]VOLTage[:DC]:RANGe
Syntax
[SENSe:]VOLTage[:DC]:RANGe <range|MIN|MAX|AUTO>
This command sets the measurement range for DC voltage measurements.
[SENSe:]VOLTage[:DC]:RANGe?
This command queries the measurement range for DC voltage measurements.
Parameter
Remarks
– Selecting a fixed range will disable autorange.
– If the input signal is greater than can be measured on the selected range
(manual range), the U2741A gives an overload indication of
+9.9E+37
from
the remote interface.
– The U2741A is set to 10 V, with autorange enabled, after a factory reset (
*RST
command)
Examples
VOLT:RANG 1
//Configures the U2741A to measure DC
voltage at 1 V range
VOLT:RANG?
//Triggers a measurement and read
Typical Response:
+1.00
See Also
[SENSe:]VOLTage[:DC]:RANGe:AUTO
Item
Type
Range of values
Default value
range
Numeric
100 mV, 1 V, 10 V, 100 V, 1 kV
MIN = 10 mV
MAX = 1 kV
10 V