Application Examples
100
Keysight U1401B User’s and Service Guide
Simulation Mode for mA Output
In simulation mode, the instrument simulates a current loop transmitter. Use this
simulation mode when an external DC 24 V or 12 V supply is in series with the
current loop being tested. Always use the special yellow test lead. Follow the
procedure below when performing mA output simulation.
1
Turn the rotary switch to any one of the
/
or
/
positions.
2
Connect the special yellow test lead between the positive output terminal of
the instrument and the positive terminal of the measurement device on the
current loop. Refer to
3
Connect the black alligator lead between the
COM
terminal of the loop source
and the negative terminal of the measurement device on the current loop.
4
Connect the red alligator lead between the negative output terminal of the
instrument and the positive terminal of the current loop source. Make sure the
polarity is correct.
5
Set the current level of the calibrator between 0 mA and 20 mA. Do not set a
negative current output value.
6
Press
OUTPUT
to output the test current.
This connection can be used for any loop voltage from 12 V to 30 V.
CAUTION
Always use the supplied special yellow test lead to perform mA simulation.
Disconnect the test lead from the current loop before turning the rotary
switch to change function or to power-off this instrument. Failure to do so
will result in a current of at least 16 mA in the 250
Ω
load connected loop.