Calibration Procedures
1
Keysight U1270 Series Service Guide
45
Adjustment procedure
1
Turn the rotary switch to the respective test function position as shown in the
adjustment input values table (
2
Unsecure the instrument to enter the adjustment mode. (See
3
The reference value of the calibration item will be shown on the display.
Figure 1-8
Reference value display
4
Configure each calibration item.
5
Use the arrow keys to select the calibration range.
6
Apply the input signal shown in the
Reference value
column of
analog bar graph displays the input reading. There is no bar graph display for
temperature adjustment.
NOTE
Review the
before
beginning the adjustment procedures.
NOTE
While in the adjustment mode, press
and
simultaneously to exit the
adjustment mode.
Shift
View
Esc
Auto
Trig
Hold