Performance Tests and Calibration
78
Keysight U1241AN User’s and Service Guide
Exiting Adjustment Mode
1
Remove all shorting plugs and connectors from the multimeter.
2
Record the new Calibration Count, see
3
Press
and
simultaneously to exit the Adjustment Mode. Power off
and on the multimeter to return to normal measurement mode and secured.
11
Press
to go to
function
10
A
0.5
A, 70 Hz
0.500 A
1
A, 70 Hz
1.000
A
10
A, 70 Hz
10.000 A
12
Turn the rotary switch to
position
Short
Dual banana plug with
copper wires short
between two terminals
SHrt
100 mV
100 mV
100.00 mV
13
Press
to go to
func tion
100 mV
3 mV, 70 Hz
3.00 mV
100 mV, 70 Hz
100.00 mV
14
Press
to go to
T1
function
[a]
K-type
0 ºC
000.0 ºC
[a] — Set the 5520A to internal reference.
— Prior to performing adjustment, connect one end of the K-type thermocouple (with miniature TC connector on both ends) to the 5520A TC
output, and the other end to a precision thermometer to verify that the source outputs the desired value. Adjust the source accordingly if
necessary.
— To perform the adjustment, connect one end of the K-type thermocouple (with miniature TC connector on both ends) to the 5520A TC
output, and the other end to the multimeter via a TC-to-banana adapter. Allow at least 1 hour for the multimeter to stabilize.
Table 5-4
Calibration adjustments (continued)
Step
Test function
Cal range
Input reference value
Cal item
U1241AN
Summary of Contents for U1241AN
Page 1: ...Keysight U1241AN Handheld Digital Multimeter User s and Service Guide ...
Page 14: ...14 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...16 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...