Performance Tests and Calibration
5
Keysight U1241AN User’s and Service Guide
71
Entering Adjustment Values
1
Press
and
to step through each character in the primary display.
2
Press
and
to change the value of the corresponding character from
digits 0 through 9.
3
Apply the respective corresponding input signal from the recommended test
4
Press
(Save) when done to start calibration.
Adjustment Considerations
1
Allow the multimeter to warm up and stabilize for five minutes before
performing the adjustments.
2
Assure that the low battery indicator does not appear during the adjustment.
Replace the batteries to avoid inaccurate readings.
3
Consider the thermal effects of the test leads connected to the calibrator and
multimeter. It is recommended to wait for one minute before start performing
the calibration.
4
During ambient temperature adjustment, ensure that the multimeter has been
turned on for at least one hour with K-type thermocouple connected between
the multimeter and calibration source.
NOTE
Ensure that the accuracy of the input signal adheres to the Valid Adjustment
Input Values in
NOTE
After each adjustment, the secondary display shows
PAS
. If the calibration fails,
the multimeter sounds a beep, and an error number is shown in the secondary
display. Calibration error messages are described in
Summary of Contents for U1241AN
Page 1: ...Keysight U1241AN Handheld Digital Multimeter User s and Service Guide ...
Page 14: ...14 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...16 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...