Performance Tests and Calibration
5
Keysight U1241AN User’s and Service Guide
65
Table 5-2
Verification tests
Step
Test Function
Range
5520A Output
Error from nominal
1 year
U1241AN
1
Turn the rotary switch to
position
1000 mV
1000.0 mV
± 1.1 mV
10 V
10.000 V
± 11 mV
100 V
100.00 V
± 110 mV
1000 V
1000.0 V
± 1.1 V
2
Press
to go to
function
1000 mV
1000.0 mV, 1 kHz
± 10.4 mV
1000.0 mV, 2.5 kHz
± 20.4 mV
1000.0 mV, 20 kHz
± 20.4 mV
10 V
10.000 V, 1 kHz
± 104 mV
10.000 V, 2.5 kHz
± 204 mV
10.000 V, 20 kHz
± 204 mV
100 V
100.00 V, 1 kHz
± 1.04 V
100.00 V, 2.5 kHz
± 2.04 V
100.00 V, 20 kHz
± 3.03 V
1000 V
1000.0 V, 1 kHz
± 10.4 V
1000.0 V, 2.5 kHz
± 30.3 V
3
Press
to go to frequency mode
100 Hz
1.000 V, 70 Hz
± 31 mHz
1000 Hz
1.000 V, 1000 Hz
± 400 mHz
10 kHz
1.000 V, 2 kHz
± 1.6 Hz
4
Turn the rotary switch to
position
Diode
1.000 V
± 5 mV