:TRIGger:ZONE<n>:STATe
Command Syntax
:TRIGger:ZONE<n>:STATe <on_off>
<n> ::= {1 | 2}
<on_off> ::= {{0 | OFF} | {1 | ON}}
<n> ::= 1-2 in NR1 format
The :TRIGger:ZONE<n>:STATe command sets the state for Zone 1 or Zone 2.
• When a zone's state is on, and the Zone Trigger feature is on (see
on page 992), that zone is actively being used to
qualify the trigger if it is not invalid or off-screen (see
• When the Zone Trigger feature is off, no zones are being used to qualify the
trigger, regardless of their individual states.
Note that :TRIGger:ZONE:STATe mimics the behavior of the
[Zone]
key on the front
panel, and :TRIGger:ZONE<n>:STATe mimics the behavior of the
Zone 1 On
and
Zone 2 On
softkeys. At least one zone's state must be on for the Zone Trigger
feature (:TRIGger:ZONE:STATe) to be on. When the states of both individual zones
are turned off, Zone Trigger is automatically turned off. In this case, when Zone
Trigger is turned back on Zone 1 is forced to on. Otherwise, if at least one zone
was on when Zone Trigger was turned off, the same configuration of individual
zone on/off states will be restored when Zone Trigger is turned back on.
Query Syntax
:TRIGger:ZONE<n>:STATe?
The :TRIGger:ZONE<n>:STATe? query returns the state of Zone 1 or Zone 2.
Return Format
<on_off><NL>
<on_off> ::= {0 | 1}
See Also
•
•
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...