892
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
28
:TIMebase Commands
:TIMebase:REFClock
Command Syntax
:TIMebase:REFClock <value>
<value> ::= {{1 | ON} | {0 | OFF}
The :TIMebase:REFClock command enables or disables the REF I/O MMCX
connector located on the front panel of the oscilloscope.
The REF I/O MMCX connector can be used as an input for the oscilloscope's
reference clock (instead of the internal 10 MHz reference), or it can be used to
output the internal 10 MHz reference clock when synchronizing multiple
instruments (see
The :TIMebase:REFClock ON command enables the REF I/O MMCX and sets the
reference signal mode to IN. The :TIMebase:REFClock OFF command disables the
REF I/O MMCX (the same as setting the reference signal mode to OFF).
Query Syntax
:TIMebase:REFClock?
The :TIMebase:REFClock? query returns the current state of the 10 MHz reference
signal mode. A "1" indicates that the REF I/O input is enabled (on), and a "0"
indicates that either the REF I/O MMCX is disabled (off) or that it is set as an
output (by the :ACQuire:RSIGnal command).
Return Format
<value><NL>
<value> ::= {0 | 1}
See Also
•
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...