SCPI Programmer's Guide
4 Sequential (Blocking) vs.
Overlapped Commands
IEEE 488.2 makes the distinction between sequential and overlapped commands
(and queries):
•
Sequential commands
also known as
blocking commands
, finish their task
before the execution of the next command starts.
These oscilloscope commands and queries are sequential (blocking):
•
:AUToscale
•
:DIGitize
•
:WMEMory<r>:SAVE <source>
•
:WAVeform:DATA?
•
Overlapped commands
run concurrently. Commands following an overlapped
command may be started before the overlapped command is completed.
Some oscilloscope commands are overlapped. For example, the oscilloscope's
save and recall commands are overlapped as well as some commands that
perform analysis.
With sequential (blocking) commands and queries, the oscilloscope is expected to
stop processing inputs, including additional remote commands and queries as
well as front panel knobs, until completed.
Pausing Execution
Between
Overlapped
Commands
With overlapped commands, you can use the *OPC? query to prevent any more
commands from being executed until the overlapped command is complete. This
may be necessary when a command that follows an overlapped command
interferes with the overlapped command's processing or analysis. For example:
:RECall:SETup "setup.scp";*OPC?;:RECall:ARBitrary "arb_wfm.csv"
You can also use the *ESR? query to look at the OPC bit (bit 0) in the Standard
Event Status Register to determine when an operation is complete.
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...