![Keysight P9241 Scpi Programmer'S Manual Download Page 376](http://html1.mh-extra.com/html/keysight/p9241/p9241_scpi-programmers-manual_1961682376.webp)
376
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
17
:FUNCtion<m> Commands
This function uses a Hilbert transform to get the real (in-phase, I) and
imaginary (quadrature, Q) parts of the input signal and then performs a
square root of the sum of the real and imaginary parts to get the
demodulated amplitude envelope waveform.
Filters operate on a single analog channel source or on a lower math function.
• Visualizations:
•
MAGNify — Operates on a single analog channel source or on a lower math
function.
•
MAXimum — This operator is like the MAXHold operator without the hold.
The maximum vertical values found at each horizontal bucket are used to
build a waveform.
•
MINimum — This operator is like the MINHold operator without the hold. The
minimum vertical values found at each horizontal bucket are used to build a
waveform.
•
PEAK — The PEAK operator is like the MAXimum operator minus the
MINimum operator. At each horizontal bucket, the minimum vertical values
found are subtracted from the maximum vertical values found to build a
waveform.
•
MAXHold — Operates on a single analog channel source or on a lower math
function. The Max Hold (or Max Envelope) operator records the maximum
vertical values found at each horizontal bucket across multiple analysis
cycles and uses those values to build a waveform.
•
MINHold — Operates on a single analog channel source or on a lower math
function. The Min Hold (or Min Envelope) operator records the minimum
vertical values found at each horizontal bucket across multiple analysis
cycles and uses those values to build a waveform.
•
TRENd — Measurement trend — Operates on a single analog channel source.
The TRENd:NMEasurement command selects the measurement whose trend
you want to measure.
Query Syntax
:FUNCtion<m>:OPERation?
The :FUNCtion<m>:OPERation? query returns the current operation for the
selected function.
Return Format
<operation><NL>
<operation> ::= {ADD | SUBT | MULT | DIV | INT | DIFF | FFT | FFTP
| SQRT | MAGN | ABS | SQU | LN | LOG | EXP | TEN | LOWP | HIGH
| AVER | SMO | ENV | LIN | MAX | MIN | PEAK | MAXH | MINH | TREN}
NOTE
If a math function is on (see :FUNCtion<m>:DISPlay), changing the operator will cause an
automatic vertical scaling of the new math function waveform. If you wait for the operation to
complete (with an *OPC? query for example), then query the math functions's scale, you can
see the vertical scaling that was automatically determined.
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...