330
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
16
:FRANalysis Commands
Introduction to
:FRANalysis
Commands
The FRANalysis subsystem controls the Frequency Response Analysis feature in
the oscilloscope.
The Frequency Response Analysis (FRA) feature controls the built-in waveform
generator to sweep a sine wave across a range of frequencies while measuring the
input to and output from a device under test (DUT). At each frequency, gain (A)
and phase are measured and plotted on a frequency response chart.
Reporting the Setup
Use :FRANalysis? to query setup information for the FRANalysis subsystem.
Return Format
The following is a sample response from the :FRANalysis? query. In this case, the
query was issued following a *RST command.
:FRAN:SOUR:INP CHAN1;OUTP CHAN2;:FRAN:FREQ:STAR +100E+00;
STOP +20.06;:FRAN:WGEN:VOLT +200.0E-03;LOAD FIFT
:FRANalysis:SOURce:IN
Put <source> (see
:FRANalysis:SOURce:IN
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels)
in NR1 format
:FRANalysis:SOURce:OU
TPut <source> (see
:FRANalysis:SOURce:OU
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels)
in NR1 format
:FRANalysis:TRACe
<selection> (see
:FRANalysis:TRACe?
<selection> ::= {NONE | ALL |
GAIN | PHASe}[,{GAIN | PHASe}]
:FRANalysis:WGEN:LOAD
<impedance> (see
:FRANalysis:WGEN:LOAD
<impedance> ::= {ONEMeg | FIFTy}
:FRANalysis:WGEN:VOLT
age
<amplitude>,[<range>]
(see
:FRANalysis:WGEN:VOLT
age? [<range>] (see
<amplitude> ::= amplitude in
volts in NR3 format
<range> ::= {F20HZ | F100HZ |
F1KHZ | F10KHZ | F100KHZ | F1MHZ
| F10MHZ | F20MHZ}
:FRANalysis:WGEN:VOLT
age:PROFile {{0 |
OFF} | {1 | ON}} (see
:FRANalysis:WGEN:VOLT
age:PROFile? (see
{0 | 1}
Table 88
:FRANalysis Commands Summary (continued)
Command
Query
Options and Query Returns
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...