:WGEN<w> Commands
31
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
1079
:WGEN<w>:VOLTage:OFFSet
Command Syntax
:WGEN<w>:VOLTage:OFFSet <offset>
<w> ::= 1 to (# WaveGen outputs) in NR1 format
<offset> ::= offset in volts in NR3 format
The :WGEN<w>:VOLTage:OFFSet command specifies the waveform's offset
voltage or the DC level. Use the :WGEN<w>:VOLTage command to specify the
amplitude.
You can also specify the amplitude and offset indirectly using the
:WGEN<w>:VOLTage:HIGH and :WGEN<w>:VOLTage:LOW commands. For
example, an amplitude of 5 V and an offset of 1 V is the same as a high-level
voltage of 4 V and a low-level voltage of -1 V.
Query Syntax
:WGEN<w>:VOLTage:OFFSet?
The :WGEN<w>:VOLTage:OFFSet? query returns the currently specified waveform
offset voltage.
Return Format
<offset><NL>
<offset> ::= offset in volts in NR3 format
See Also
•
"Introduction to :WGEN<w> Commands"
•
•
•
•
NOTE
When the amplitude is below 40 mV, the offset is limited to ±500 mV.
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...