:WGEN<w> Commands
31
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
1059
:WGEN<w>:MODulation:FM:DEViation
Command Syntax
:WGEN<w>:MODulation:FM:DEViation <frequency>
<w> ::= 1 in NR1 format
<frequency> ::= frequency deviation in Hz in NR3 format
The :WGEN<w>:MODulation:FM:DEViation command specifies the frequency
deviation from the original carrier signal frequency.
When the modulating signal is at its maximum amplitude, the output frequency is
the carrier signal frequency plus the deviation amount, and when the modulating
signal is at its minimum amplitude, the output frequency is the carrier signal
frequency minus the deviation amount.
The frequency deviation cannot be greater than the original carrier signal
frequency.
Also, the sum of the original carrier signal frequency and the frequency deviation
must be less than or equal to the maximum frequency for the selected waveform
generator function plus 100 kHz.
Query Syntax
:WGEN<w>:MODulation:FM:DEViation?
The :WGEN<w>:MODulation:FM:DEViation? query returns the frequency deviation
setting.
Return Format
<frequency><NL>
<frequency> ::= frequency deviation in Hz in NR3 format
See Also
•
":WGEN<w>:MODulation:AM:DEPTh"
•
":WGEN<w>:MODulation:AM:FREQuency"
•
":WGEN<w>:MODulation:FM:FREQuency"
•
":WGEN<w>:MODulation:FSKey:FREQuency"
•
":WGEN<w>:MODulation:FSKey:RATE"
•
":WGEN<w>:MODulation:FUNCtion"
•
":WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry"
•
•
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...