Time Domain – Option 010
Time Domain (Transform) Settings
Stimulus (Mode)
— There are three variations on how the Time Domain transform algorithm is
applied to the frequency domain measurement. Each method has a unique
application.
—
Lowpass Impulse
- Highest resolution. Most useful for seeing small
responses in devices that pass low frequencies, such as cables.
—
Lowpass Step
- Easiest to identify inductive and capacitive discontinuities
in devices that pass low frequencies such as cables.
In both Lowpass modes, frequencies down to DC and negative frequencies are
extrapolated. Therefore, when either Lowpass mode is selected,
Set Freq.Low
Pass
is automatically applied to adjust the start frequency. Learn more about
Set Frequency Lowpass below.
When the Start Frequency or resolution is changed AFTER selecting a Lowpass
mode, then Stimulus mode is set to Band Pass and Transform is disabled.
—
Band pass
- Easiest method - can be used with any frequency sweep. Most
useful for measuring band limited devices such as filters and DC blocked
cables. This mode does NOT show capacitive and inductive reactance. For
the same frequency span and number of points, band pass mode has twice
the impulse width, which hides closely spaced responses degrading the
response resolution.
Set Frequency Lowpass
USE ONLY IN LOW PASS MODES. The Low Pass Start frequency is adjusted so
that the Low Pass Stop frequency is at harmonics of the Low Pass Start
frequency. Start frequency is computed by the following formula:
Low Pass
Start Frequency = Stop Frequency / Number of points
.
Start/Stop Time
The following settings adjust the
display resolution
, allowing you to zoom IN
or OUT on a response. When the start or stop time is updated, they may be
automatically adjusted to limit the display to one alias-free response on either
side of zero time.
How to set Start and Stop time
— Press
Measure Setup 4
— Then
Transform
— Then
Transform Start Stop
This setting is made automatically when either Lowpass mode is selected.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...