64 Keysight N9923-90001 User’s Guide
Time Domain – Option 010
Time Domain (Transform) Settings
Time Domain (Transform) Settings
You can set and view most of the Time Domain settings on the Transform
Settings table.
Transform Settings Table
The following settings (from Transform Stimulus Settings to Data Processing)
are listed in the order they appear on the Transform Settings menu selection.
How to make settings on the Transform settings table
— Press
Meas Setup 4
.
— Then
Transform
— Then
Transform Settings
.
— Press
Next Page
and
Previous Page
to view all settings.
— To change a setting:
—Use the
arrows or rotary knob to highlight a setting.
—Numeric settings can be changed by pressing numbers using the
numeric keypad. Then press Enter or select a suffix if available.
—Some settings allow you to press
Edit
to toggle the value. Other
settings require that you then press a softkey to change the value.
—When finished changing a value, press
Done Edit
.
— Press
Dock Window
to relocate the Settings table to a position relative to
the trace window. The Dock Window setting persists through a Preset.
Choose from the following:
—
Full (Default setting)
Only the Settings table is shown on the screen.
The trace window is temporarily not shown.
—
Left
The Settings table is shown to the left of the trace window.
—
Bottom
The Settings table is shown below the trace window.
— When finished changing ALL settings, press
Done
to save your settings.
Frequency Range and Points
Like CAT mode, all Time Domain measurements are made in the frequency
domain and, using Inverse Fourier Transform (IFT), time is calculated. Select
the frequency range from which Time Domain measurements are calculated.
Increasing the data points will improve measurement resolution. However,
more data points will usually result in slower sweep updates.
Set frequency range and points before selecting Stimulus Mode.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...