Keysight N9923-90001 User’s Guide
253
Safety Considerations
Battery Protective Functions
The protective functions can be divided between two categories: active and
passive. Active protection refers to the type of protection that depends on at
least two or more protection devices working together to enable the protection
Passive protection refers to the type of protection that is always enabled
independent of any other protection device.
Active Protection
Primary Overcharge Voltage:
This prevents the battery from being charged if
the voltage across any cell exceeds approximately 4.35 V per cell. Once the
overcharge protection is tripped, the voltage across each cell must drop below
approximately 4.15 V to reset the protection and permit charging.
Secondary Overcharge Voltage:
This is a one-time, permanent protection that
is triggered when the voltage across any cell exceeds approximately 4.45 V.
Once this protection is tripped, the battery can no longer be used and must be
replaced.
Overdischarge Voltage:
This prevents the battery from discharging if the
voltage across any cell drops below approximately 2.6 V. Once the
overdischarge voltage protection is tripped, the voltage across each cell must
be charged to approximately 3 V to reset the protection and permit
discharging.
Primary Overcharge Current:
This prevents the battery from being charged if
the average charging current reaches or exceeds 3.5A for a time period of 120
seconds or more.
Secondary Overcharge Current:
This prevents the battery from being charged
if the charging current reaches or exceeds 4A+/–20% for a time period of 20
milliseconds or more.
Primary Overdischarge Current:
This prevents the battery from being
discharged if the average current out of the battery reaches or exceeds 5A for a
time period of 120 seconds or more. This protection can be reset by removing
the load.
Secondary Overdischarge Current:
This prevents the battery from being
discharged if the current out of the battery reaches or exceeds 6A for a time
period of 20 milliseconds or more. This protection can be reset by removing the
load.
Short Circuit Protection:
This prevents the battery from being charged or
discharged and protects against damage or lost data if the current in or out of
the battery reaches or exceeds 12.5A for a time period of 183 microseconds or
more. This protection can be reset by removing the load.
Overtemperature Charging:
The system microcontroller prevents the battery
from being charged if the cell temperature exceeds 46C. Once the
overtemperature charging protection is tripped, the cell temperature has to
drop to or below 45C to reset the protection and permit charging.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...