152 Keysight N9923-90001 User’s Guide
Pulse Measurements Mode – Option 330
Pulse Measurement Settings
How to set Single or Continuous
— Press
Sweep 3
— Then choose one of the following:
—
Single
Automatically sets Continuous OFF and causes the FieldFox
to query the USB Power Sensor ONCE, then hold for the next Single
key press. Hold is annotated in the upper left corner of the display.
—
Continuous
Queries the USB Power Sensor whenever a valid trigger
signal is present. This is the typical setting when battery power is not
critical.
You can also use
Run / Hold +/-
to toggle between Single and Continuous.
Resolution
Resolution is the number of individual acquisitions that occur in the USB
Power Sensor. This data is read by the FieldFox and plotted across the X-axis
to form a trace. More data points increases measurement resolution. However,
more data points requires more time to acquire, query, and plot.
How to set Resolution
— Press
Sweep 3
— Then
Resolution
— Then choose from the following:
—
Low
240 data points.
—
Medium
1000 data points.
—
High
Calculated value - approximately 8000 data points.
Triggering
For pulse measurements, a trigger signal refers to the event that causes the
USB Peak Power Sensor to initiate measurement acquisition. It does NOT refer
to triggering the FieldFox.
Trigger settings can be used with ALL pulse measurements unless stated
otherwise.
A
Waiting for Trigger
message on the FieldFox screen is posted on behalf of
the USB Power Sensor when it is waiting for a pulse signal at the RF Input
(Internal) or a TTL signal at the USB Sensor External Trigger input.
How to make Trigger settings
— Press
Sweep 3
— Then
Trigger
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...