128 Keysight N9923-90001 User’s Guide
Power Meter (USB) Mode
USB Power Meter Settings
USB Power Meter Settings
Select Power Meter Mode before making any setting in this section.
How to select Power Meter Mode
— Press
Mode_
.
— Then
Power Meter (USB Sensor)
.
Average / Peak
— Press
Measure 1
— Then choose from the following:
—
Average
Measures and reports the average of the power
measurements.
—
Peak
Available ONLY on the U202x power sensors. Measures and
reports peak power detected at sensor input.
Zeroing
The Keysight USB Power Sensors perform Internal Zeroing automatically.
Because Keysight USB Power Sensors have an internal switch, Internal Zeroing
does NOT require that the power source be turned OFF.
For highest measurement accuracy, when measuring power levels below –30
dBm, External Zeroing should be performed.
During External Zeroing, the power source MUST be turned OFF or the power
sensor disconnected from the power source (device under test). Power is OFF
when power sensor measurements are below –60 dBm. When power higher
than this is detected at the sensor, a warning message is shown on the
FieldFox and External Zeroing will not proceed.
How to perform External Zeroing
— Press
Cal 5
.
— Then
External Zero
.
While either Internal or External Zeroing is being performed, “Zeroing...” is
shown on the FieldFox screen. While this message is present, do NOT make
any changes to the FieldFox.
External Zeroing is NOT allowed on the U2020 X-series Power Sensors.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...