Keysight N9923-90001 User’s Guide
113
Calibration for NA, CAT, and VVM Modes
Normalize
uses a THRU standard or cable between port 1 and port 2 to cal an
S21 and S12 Transmission measurement (NA Mode) and a 2-port Insertion
Loss measurement (CAT Mode). In VVM Mode, this is performed using Zero.
How to perform a Response Cal
— Select the measurements to be calibrated. See the relevant Mode (NA, CAT,
or VVM) for measurement selections.
— Press
Cal 5
then
Response Cal
— For 1-port measurements:
—Select either
Open Response
or
Short Response
for the port to be
calibrated. The availability of this calibration on Port 2 may require an
option.
—Connect an OPEN or SHORT standard to the specified port and press
Measure
— For 2-port measurements:
—Select
Normalization
—Connect a short, high-quality, phase stable cable between the
FieldFox port 1 and port 2 connectors, then press
Measure
— Press
Finish
.
— Connect the DUT.
CAL ON U
is shown on the screen when a User Cal (Ex: Response Cal) is
correcting ONLY the appropriate measurement. For example, when an Open
Response Cal on Port 2 is performed,
CAL ON U
is shown for an S22
measurement only.
With a Normalization Cal, all subsequent insertion loss measurements are
made relative to the insertion loss of the cable used as the THRU standard.
For example, if you use a cable with 1 dB of loss, then after Normalization,
the display will show 0 dB of loss with this cable in place. Therefore, for
highest accuracy, when measuring the DUT also attach the cable that was
used in the normalization cal.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...