108 Keysight N9923-90001 User’s Guide
Calibration for NA, CAT, and VVM Modes
—When your DUT has high return loss such as the reject band of a filter.
Then measuring the load improves accuracy. Even with the optional
Load, phase accuracy begins to degrade when the return loss is
greater than about 20dB.
— Press
Skip Step
in all other cases and when making DTF measurements.
Measuring a load would do little to improve accuracy.
Connect THRU (Not used for 1-port cals)
— Connect port 1 to port 2, either by mating the connectors that attach to the
DUT (Flush THRU), or by using a short jumper cable. Learn more about
Flush THRU in
“Calibration Type” on page 114
. Then press
Measure
.
— Press
Finish
.
CAL ON Q
is shown on the screen for all displayed measurements that are
corrected with QuickCal.
Using QuickCal with a Long Jumper Cable
QuickCal requires a sufficient frequency span to adequately characterize the
jumper cable (the cable between the instrument port and the DUT). If a ‘Cal
Failed’ error appears and you connected the optional Load standard, then
increase the frequency span of the measurement until the error no longer
appears. After the calibration, you can then decrease the frequency span as
needed and rely on interpolation. Learn more about Interpolation in
QuickCal is also limited in how much jumper cable loss it can tolerate. If
QuickCal without the Load is failing, then measure the Load during the
QuickCal to see if that resolves the issue.
Mechanical Cal
Mechanical Calibration is performed using discrete standards from a Cal Kit.
Several Cal Kit definitions are built into the FieldFox. To learn about Cal Kit
definitions, refer to the Application Note, “Specifying Calibration Standards
and Kits for Keysight Vector Network Analyzers,” available online at
http://literature.cdn.keysight.com/litweb/pdf/5989-4840EN.pdf
Mechanical Cals are extremely accurate when performed using the correct Cal
Kits with standards that are clean and in good repair, and when using correct
connection procedures.
How to perform a Mechanical Cal
— Disconnect the DUT from the FieldFox.
Visit
www.keysight.com/find/fieldfoxsupport
to see a complete list of
supported Cal Kits. Also at this website, download Data-Link software that
allows you to edit Cal Kit definitions or add a new Cal Kit.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...