106 Keysight N9923-90001 User’s Guide
Calibration for NA, CAT, and VVM Modes
QuickCal (Option 112)
QuickCal is ideal for a quick and simple calibration of the measurement when
using a jumper cable or adapter to connect the DUT to the FieldFox test ports.
QuickCal corrects for phase shift, time delay, and loss of adapters or jumper
cables.
QuickCal does not require a specific Cal Kit and is a very simple calibration. If
the highest accuracy is required, a mechanical calibration with a high quality
calibration kit is recommended.
QuickCal is designed and most accurate when calibrating for DUTs fitted with
Type-N and 7/16
connectors.
If the DUT has different connectors, such as 3.5 mm (male) or SMA male
connectors, QuickCal can still be used by selecting 'Other' connector.
However, the accuracy of QuickCal with ‘Other’ connectors is not specified.
If the DUT has 3.5 mm-female or SMA-female connectors, QuickCal is NOT
recommended. For accurate and repeatable measurements of devices with
female 3.5 mm or SMA connectors, use mechanical calibration.
Important QuickCal Notes
— 7/16 connectors can be used with QuickCal ONLY up to 6 GHz.
— Type-N 75 Ohm can be used with QuickCal ONLY up to 3 GHz.
— QuickCal cannot be used with waveguides, or if the test fixture or test setup
(before DUT) includes an attenuator.
How to perform a QuickCal
— In CAT, NA, or VVM Mode, press
Cal 5
— Then
QuickCal
—Press
Change DUT Connectors
to select the correct connector types
that are on your DUT. Although there is no Cal Kit required, selecting
the connector type (limited to Type-N and 7/16) and gender from the
list provides an accurate OPEN model which means a more accurate
calibration. If the DUT connector type is not listed, select
Unknown/Other.
Source Unleveled errors - During calibration, the frequency range of the
measurement MAY be extended to provide maximum flexibility. During the
calibration, the output power may become unleveled at the added high
frequencies. You can ignore the “Source Unleveled” error, or to avoid the
error, select either High power or -15 dBm before calibrating. Learn more
about setting Output Power in
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...