Keysight N9038A MXE Service Guide 667
Tests
Before Performing a Functional Test
Before Performing a Functional Test
1.
Ensure that you have the proper test equipment.
2.
Switch on the unit under test (UUT) and let it warm up (in accordance with
warm-up requirements in the instrument specifications).
3.
Allow sufficient warm-up time for the required test equipment (refer to
individual instrument documentation for warm-up specifications).
4.
Ensure that the analyzer’s frequency reference is set to Internal:
a.
Press the
Input/Output
, More,
Freq Ref In
keys.
b.
If the
Freq Ref In
softkey does
not
show Internal, press the
Freq Ref
In
softkey and select Internal.
5.
Following instrument warm-up, perform the auto align routine:
Press
System
,
Alignments
,
Align Now, All
.
6.
Run the EMI receiver internal RF preselector alignments by pressing
System
,
Alignments
,
More
,
RF Preselector
,
Align Now
,
20 Hz to 3.6 GHz
.
Functional test accuracy depends on the precision of the test equipment used. Ensure that all of
the test equipment is calibrated before running a functional test.
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...