Post-Repair Procedures
Post-Repair Procedures
FPGA Synchronization
There is FPGA (Field-Programmable Gate Array) program code on many
different assemblies inside of the instrument, and all of these must be of a
compatible version with the instrument software revision. In order to ensure
that this requirement is met when either an assembly is replaced that contains
FPGA code on it, or if the A5 Disk Drive is replaced with a drive that has a
different version of software than the previous drive, you will need to take
certain steps to ensure that the instrument operates properly.
Updating the Instrument FPGA Code
All instruments have a utility included in them that will update the FPGA code
on all of the different assemblies in the instrument that have it. When this
utility is run it will detect the assemblies that need to be updated and will
update them to the correct revision.
1.
Close the instrument application software by pressing
File
,
Exit
,
Enter
using the front panel keys.
2.
Connect a USB mouse to one of the instrument front panel USB ports.
3.
Using the mouse go to Start, Computer and navigate to the following
folder:
C:\Program Files\Agilent\SignalAnalysis\Physics
4.
In this folder find and execute the file named:
FPGA_Prog.bat
5.
The FPGA Programming Utility will start and a window as shown in
will appear.
Even if an instrument appears to work properly when an assembly with FPGA code is replaced,
that does not mean that the FPGA code is completely compatible with the software version
installed in the instrument. There could be incompatibilities that will only be seen under certain
measurement conditions.
Once you start the FPGA programming process you MUST NOT interrupt the process for any
reason. This would include turning the instrument off or unplugging the power cord to the
instrument. Doing so will result in an inoperative instrument, requiring the affected assembly to
be replaced.
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...