Keysight N9038A MXE Service Guide 533
Assembly Replacement Procedures
Instrument Outer Case
Figure 18-3
Instrument Outer Case Removal
Replacement
1.
Disconnect the instrument from ac power.
2.
Slide the instrument cover back onto the deck from the rear. The seam on
the cover should be on the bottom. Be sure the cover seats into the gasket
groove in the Front Frame Assembly.
3.
Replace the four rear feet to the rear of the instrument. Torque the rear
feet screws (
washers) to 21 inch pounds.
4.
Replace the bottom feet by sliding into place until they snap in. Install the
locks by pressing in flat.
5.
Replace the handle straps on both sides of the instrument. Torque the
handle strap screws to 21 inch pounds.
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...