DRAM
128 MB volatile working memory. This contains data for
current and recent calibrating and measuring.
Clearing User Data
Clearing the DRAM
All data in the DRAM is cleared by powering the instrument off.
Clearing the Flash ROM
The user data can be cleared from the Flash ROM by:
1
Reset the instrument to factory default.
a
Send the command:
SYSTem:PRESet
OR
b
Press the LAN Reset button on the front panel briefly (<1 s).
2
Clear the instrument log.
a
Send the special command
DIAG:SYST:FLAS:LOGF:ERAS
The contents of the log can be checked with the query READ:LOGF?.
3
Clear the Ethernet settings
a
Connect to the instrument by some means other than the LAN,
using VISA Assistant (for example over USB). Send the command:
SYSTem:COMMunicate:ETHernet:RESet
OR
b
With the instrument connected over the LAN (the LAN LED is
green), press and hold the LAN Reset button for more than 3s.
In either case, the instrument reboots.
Maintenance and Troubleshooting
6
Keysight N774x Series
99
Summary of Contents for N7744A
Page 2: ......
Page 74: ...Specifications and Regulations Compliance 4 Keysight N774x Series 73 ...
Page 76: ...General Characteristics Specifications and Regulations Compliance 4 Keysight N774x Series 75 ...
Page 101: ...6 Maintenance and Troubleshooting 100 Keysight N774x Series ...
Page 104: ......