Sink Calibration Procedures
Sensitivity Test
Purpose and Method
This test characterizes that the DUT can still work with the minimum
differential voltage.
The procedures starts at the maximum output voltage and performs the
Link Training. If the device passes, the generator output is decreased and
tested again. This is repeated until the device fails because the bits are not
detected properly anymore owing to the low voltage level applied.
The table also shows the eye opening corresponding to the set amplitude.
This value is derived from the
Connection Diagram
Figure 41
Connection Diagram for Receiver Tests
Parameters in Expert Mode
The parameters used in expert mode for this calibration are:
• Bit Error Rate Testing