Installation Note N5242-90123
38
Installation Procedure for the Upgrade
Step 31. Remove Option 222 License
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must be running.
— A keyboard and mouse must be connected to the network analyzer.
Option 222 License Removal Procedure
1.
To start the Keysight License Manager, press Start > Keysight License Manager > Keysight
License Manager. A Keysight License Manager dialog box will appear.
2.
Right click the on the desired option and click
Delete
.
3.
In the Select Desired Option list, click
222
.
4.
In the Keysight License Manager dialog box that appears, press or click
Yes
to confirm delete.
5.
A message displays stating that the option removal was successful.
6.
Restart the PNA Analyzer application: Press
File
>
Exit
.
7.
In the Exit NA Application dialog box that opens, press
OK
.
Step 32.Enable Option 422
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must be running.
— A keyboard and mouse must be connected to the network analyzer.
— Refer to the license message you received from Keysight: Verify that the analyzer’s model and
serial numbers match those on the license message you received from Keysight.
— Obtain your license key file(s) for installation of this upgrade by following the instructions on the
supplied Software Entitlement Certificate.
For this step, you will need a USB flash drive.