Keysight N5224B/25B Service Guide 7-59
Repair and Replacement Procedures
Removing and Replacing the A70 or A75 Low Frequency Extension (LFE) Board
Removing and Replacing the A70 or A75 Low Frequency Extension
(LFE) Board
Tools Required
— T-10 TORX driver (torque to 9 in-lb or 1.02 N.m)
— T-20 TORX driver (torque to 21 in-lb or 2.38 N.m)
— 5/16 inch open-end torque wrench (torque to 10 in-lb or 1.13 N.m)
— ESD grounding wrist strap
Removal Procedure
for this procedure.
1.
Disconnect the power cord.
2.
Remove the outer cover. Refer to
“Removing the Covers” on page 7-7
3.
Position the analyzer bottom side up.
4.
Remove the A23 test set motherboard. Refer to