Using the Probes
1
N2795/6A User’s Guide
19
Dynamic Range and Offset Voltage Limitations
With 16V
pp
(±8V) of dynamic range and ±8V (N2795A) or ±12V
(N2796A) of offset range, the probes have a wide measuring range
of ±20V, which allows them to be used for a wide variety of
applications. For example, if you wanted to measure small AC noise
riding on top of a +10V DC signal, you could use DC offset to
position the signal at the center of the screen to keep the probe
within its dynamic range. You could then zoom in to see the detail
at a higher resolution. Trying to measure a signal out of the
dynamic range may result in a clipped or distorted waveform.
Figure 10
Dynamic Range and Offset Limits
Summary of Contents for N2795/6A
Page 1: ...Keysight N2795 6A Single Ended Probes User s Guide...
Page 4: ...4 N2795 6A User s Guide Contents...
Page 29: ...Characteristics and Specifications 2 N2795 6A User s Guide 29 Dimensions...
Page 30: ...30 N2795 6A User s Guide 2 Characteristics and Specifications...
Page 34: ...34 N2795 6A User s Guide 3 Performance Data Plots...
Page 48: ...48 N2795 6A User s Guide 4 Performance Verification...
Page 50: ...Index 50 N2795 6A User s Guide...