![Keysight M9202A Startup Manual Download Page 19](http://html1.mh-extra.com/html/keysight/m9202a/m9202a_startup-manual_1961302019.webp)
Step 5: Verify Operation of the M9202A Module
Keysight M9202A PXIe Digitizer, Startup Guide
19
Conducting a Self-Test
The purpose of this self-test is to verify that the unit functions correctly, in that it can
self-calibrate and communicate with the processor controlling it.
Start the SFP 1 from Window Start Menu > Agilent > MD1 Digitizer > SFP 1.11.
When the SFP 1 is launched it will initiate a self-calibration which might take a little
over 1 minute.
Verify that no error is shown in the lower left hand log area. The image below shows
the desired result.
Soft Front Panel (SFP 1) showing the result of the Self-Test
Perform a Verification of the M9202A (optional)
Requirements for Verification
The M9202A is verified by using it to trigger on and visualize a signal from a Function
Generator. The trigger must be stable and the signal frequency and amplitude must
correspond to that set on the generator.