Test 1: DC Volts Verification Tests
18
Keysight M9182A/M9183A Calibration and Verification Procedure
Test 1: DC Volts Verification Tests
1
Make sure you have read
2
Connect the calibrator to the input terminals of the DMM, as shown below.
3
Program a null reading and make all DCV measurement in the following table
“nulled” to this measurement. See
4
On the M9182A/83A DMM, select each function and range in the order shown
in thefollowing table. Using the calibrator, provide the input shown in the same
table.
5
Make a DC voltage measurement on the M9182A/83A DMM and return the
result. Compare measurement results to the appropriate test limits shown in
the table. Be certain to allow for appropriate source settling when using the
Fluke 5720A.