Keysight M809256PB OIF CEI-56G Rx Test Automation Application User Guide
3
Preparing to take Measurements
Calibrating the Instruments
If you haven't already calibrated the FlexDCA oscilloscope, Keysight
recommends calibrating the oscilloscope before performing calibrations or
tests using the OIF CEI-56G Rx Test Automation Application.
1 Disconnect all the cables that may be connected to the FlexDCA
oscilloscope.
2 From the main menu of the FlexDCA N1000-Series System Software,
select
Tools
>
Calibrations...
.
3 Follow the on-screen instructions to perform calibrations.
If the calibrations are not performed, warning messages are logged for
each calibration performed from the M8070B software.
To know how to perform calibrations on the oscilloscope, refer to the
DCA-X, DCA-M and FlexDCA Help v6.40 User's Guide
for more details.